1. AUTOMATIC
CONTROL SYSTEM:
The
Machine is equipped with automatic control system by Pre-programmed PLC.
|
2.
OPERATION SYSTEM:
The Machine
is capable of running in automatic mode via 4.7” colour touch
screen (HMI).
|
3. PRE-PROGAMMED
CYCLES:
The sterilizer has equipped 5 pre-programmed cycles
as mentioned below:
a)
Liquid Cycle
b)
Packs Cycle
c)
Flexible Cycle
d)
Wrapped
Cycle
e)
Unwrapped
Cycle
|
4. USER
CUSTOMIZABLE CYCLE:
The machine
having user variable
cycle will allow user to customize the temperature settings, pressure settings and vacuum
pulse settings.
|
5. PASSWORD
PROTECTION:
a)
The sterilizer settings are password protected.
b) The date and time settings are password protected, and User is able to update date and time settings on the screen
for better record
keeping.
|
6. PRINTER:
a)
The user is record the progress of the sterilization
cycle in real time therefore online real time printing facility will be
provided.
b)
The user can print
the following details:
· Date
and time of the cycle.
· Status
of the cycle.
· Temperature
at drain and chamber pressure at every event change and during sterile hold
it can be printed at desired interval, minimum of 1 minute.
· All
undesired situation (faults) are to be printed at the time of fault
generation.
Optional:- Printer will
be provided at extra cost with this model.
|
7. CYCLE DATA
STORAGE FACILITY:
Historical
cycle data gets stored in the memory of sterilizer which can be easily
retrieved at will through USB.
|
8. DOOR:
· The
Sterilizer has manual sliding single door.
· Door
frame is strong & robust enough to ensure smooth door movement &
consistency in alignment.
· Door
movement is smooth to ensure proper door alignment over a long period of
time.
|
9. QUICK HOT AIR DRYING SYSTEM:
Specially
designed drying valves have been provided to ensure each load comes out
completely dry.
|
11. STATE OF THE ART
– VACUUM SYSTEM:
Heavy
Duty Branded- Diaphragm type vacuum
pump is provided to remove the steam from the chamber.
|
12. AUTOMATIC TEMPERATURE
DETECTION SYSTEM:
The machine is equipped with temperature
detection system and the user will be notified in case the machine fails to
reach the set temperature or if the temperature over shoots.
|
13. AUTOMATIC LOW
WATER CUT OFF SYSTEM:
The
machine is equipped with automatic low water cut-off system. When the water
level decreases below a threshold then heating elements will automatically
turn off. The user will be notified in case the water level is not optimum
for regular machine operation.
|
14. AUTOMATIC PRESSURE CONTROL SYSTEM:
The sterilizer has automatic pressure control switch
which automatically maintain the desired pressure and restricts any abnormal/
unsafe rise of pressure in the sterilizer.
|
16. AIRTIGHT SEALING:
Heavy
Duty branded silicon gasket will be provided for airtight seal.
|
17. SOLENOID VALVE:
The sterilizer is equipped with branded solenoid
valves for automatic process.
|
18. ANALOGUE GAUGES:
a) The
Sterilizer is fitted with branded NABH calibrated analogue chamber pressure (Compound) gauge to measure the
pressure and vacuum of chamber.
b) The
Sterilizer is fitted with branded NABH calibrated analogue jacket pressure gauge to measure the jacket & boiler
pressure.
|
19. DIGITAL
TEMPERATURE INDICATOR:
The Sterilizer is fitted with digital
temperature indicator to indicate
the chamber temperature in
manual mode.
|
20. EUROPEAN STYLE
OUTER COVER:
The sterilizer has fully
covered design with panelling back and both the sides RHS & LHS.
|
21. SAFETY FEATURES:
21.1 EMERGENCY STOP:
An Easily
Accessible emergency stop is provided on main control
panel door.
21.2 POWER FAILURE:
Equipment settings will not get disturbed due
to power failure and will not restart without human intervention.
21.3 FAILURE MODE
DETECTION:
Equipment indicate fault through audio / visual
signal with details
on Control panel / PLC (Failure mode detection).
21.4 HYDRAULIC TESTING:
Complete sterilizer including chamber, jacket, and
piping are hydraulically tested at 2 times the working pressure.
21.5 INSULATION:
The sterilizer is completely insulated to prevent any
burns caused due to accidental contact to sterilizer body.
|